2000A Hyper-Link CPG for 81nt Oligonucleotide Synthesis Test Report

Introduction

This report presents a comprehensive performance evaluation and comparative analysis across four distinct CPG solid support samples used for oligonucleotide synthesis.

  • Conventional Functionalized CPG Group (Samples 0309-1 & 0309-2): These samples utilize standard, high-quality commercial CPG1000Å solid supports, sourced from a leading industry provider (brand name withheld for confidentiality). These represent the current market benchmark for high-performance synthesis.
  • Poresyn Co-polymer Coated CPG Group (Samples 0309-3 & 0309-4): These samples utilize Poresyn Solutions‘ proprietary 2000Å Co-polymer Coated CPG (PS+CPG Hyper-Link Architecture) with pre-installed Unylinker. This next-generation carrier features a unique crosslinked structure of PS resin and CPG, engineered to provide the mechanical stability of glass with the high loading capacity of polymers.

Through the data provided in this report, a clear performance gap is observed between conventional CPG and Poresyn’s new polystyrene+CPG architecture.

Synthesis Test Report – Poresyn Solutions

Synthesis Performance Analysis

Poresyn Solutions | Advanced Biomaterials Division
Internal Lab Report

Molecular Information

Synthesized Sequence: 81nt GCCTXXXXX….TTGT
Molecular Weight: M.W 24837.16
Synthesis Scale: 200nmol – 1.0μmol
Test Date: 2026.03.09 – 03.11

Process & Analysis

Cleavage/Deprotection: Liquid Phase AMA
Conditions: 65°C | 2 Hours | 1ml Vol.
A260 Determination: Bioteck (100x Dilution)
Purity Method: Agilent 1260 Infinity II

II. Quantitative Result Table

Sample IDSolid Support ArchitectureA260Total ODNet NMOLYield %Purity %
0309-1Conventional Functionalized CPG 1000Å0.656581.940.95%22.96%
0309-2Conventional Functionalized CPG 1000Å2.65265333.933.39%19.53%
0309-3PORESYN 2000Å (DMT-OFF)3.30330415.841.58%43.80%
0309-4PORESYN 2000Å (DMT-ON)3.65365459.945.99%41.87%

Performance Conclusion

  • Superior Crude Purity: Poresyn Hyper-Link 2000Å support demonstrated a purity increase of 110% to 120% over the 1000Å conventional functionalized CPGs, confirming excellent chemical inertness for mid-length oligos.
  • Enhanced Molar Output: At the 1.0μmol scale, Poresyn achieved up to 459.9 NMOL, significantly outperforming the industry conventionalized CPG by over 37%.
  • Thermal & Chemical Stability: Consistent data across DMT-ON/OFF states confirms that the co-polymer coating remains stable under aggressive AMA deprotection conditions (65°C).

Poresyn Solutions | www.poresynsolutions.com

This report is generated for performance validation of Hyper-Link series solid supports.

HPLC Data – 0309-1

HPLC Injection Parameters Table
HPLC Acquisition & Processing Parameters
Data File0309-1 20260310 183312.dx
Sequence NameSingleSample
Project NameTest
Sample Name0309-1
Operatoradmin
InstrumentLC1260II
Injection Date2026-03-11 12:26:18+08:00
Injection Volume10.000 μL
LocationP2-D5
Acquisition Method45-100nt DNA.amx
TypeSample
Processing MethodDNA-100NT.pmx
Sample Amount0.00
Manual ModificationNone


HPLC Peak Analysis Report, Signal: VWD1A, Wavelength=254 nm

Ret. Time [min]TypeWidth [min]AreaHeightArea %
9.654BV0.2730.613.160.21
9.844VV0.1828.593.310.20
10.032VB0.2329.403.660.20
10.618BB0.29117.2516.380.82
10.922BV0.2120.163.110.14
11.036VB0.149.891.960.07
11.246BV0.27117.9417.970.82
11.502VB0.1722.424.600.16
11.673BV0.1798.6217.570.69
11.856VV0.22239.4926.771.67
12.044VV0.14159.6823.771.11
12.186VV0.17210.3628.131.47
12.363VB0.35262.2024.411.83
12.770BV0.31230.1824.331.60
12.993VB0.1314.972.800.10
13.219BV0.2272.208.590.50
13.535VB0.38120.017.830.84
14.057BV0.53499.5824.823.48
14.357VB0.26105.9411.920.74
14.713BV0.34179.4713.071.25
15.172VV0.47561.6924.343.92
15.419VV0.24366.8927.372.56
15.878VV0.723292.93195.7722.96
17.266VB2.467553.61111.9852.66
Total14344.11100.00

HPLC Data – 0309-2

HPLC Injection Parameters – Sample 0309-2
HPLC Acquisition & Processing Parameters Sample: 0309-2
Data File0309-2 20260310 183333.dx
Sequence NameSingleSample
Project NameTest
Sample Name0309-2
Operatoradmin
InstrumentLC1260II
Injection Date2026-03-11 12:52:03+08:00
Injection Volume2.000 μL
LocationP2-D6
Acquisition Method45-100nt DNA.amx
TypeSample
Processing MethodDNA-100NT.pmx
Sample Amount0.00
Manual ModificationNone

HPLC Peak Analysis Report – Sample 0309-2,Signal: VWD1A, Wavelength=254 nm

Ret. Time [min]TypeWidth [min]AreaHeightArea %
9.418BB0.3224.051.950.34
9.619BV0.1916.772.480.23
9.804VB0.2928.303.570.40
10.488BB0.3083.2510.931.16
10.802BV0.2018.873.280.26
10.926VB0.1718.283.200.26
11.161BV0.2582.9112.291.16
11.404VB0.2030.275.430.42
11.606BB0.1745.139.020.63
11.823BV0.24179.4619.502.51
11.992VV0.1499.0815.351.38
12.152VV0.18137.5217.141.92
12.352VB0.37194.6114.972.72
12.762BB0.48227.7518.683.18
13.254BV0.1919.802.350.28
13.459VB0.2524.963.010.35
14.175BV0.55260.7813.783.64
14.515VB0.2949.374.400.69
14.900BV0.2650.985.190.71
15.332VV0.46128.124.481.79
15.983BV0.711398.5690.7919.53
16.454VV0.20192.4617.762.69
17.248VBA2.863850.6959.6453.77
Totals7161.99100.00

HPLC Data – 0309-3 (PoresynSolutions’ 2000A Co-Polymer Coated CPG DMT-OFF)

HPLC Injection Parameters – Sample 0309-3
HPLC Acquisition & Processing Parameters Sample: 0309-3
Data File0309-3 20260310 183342.dx
Sequence NameSingleSample
Project NameTest
Sample Name0309-3
Operatoradmin
InstrumentLC1260II
Injection Date2026-03-11 13:17:47+08:00
Injection Volume2.000 μL
LocationP2-D7
Acquisition Method45-100nt DNA.amx
TypeSample
Processing MethodDNA-100NT.pmx
Sample Amount0.00
Manual ModificationManual Integration (手动积分)



HPLC Peak Analysis Report – Sample 0309-3

HPLC Peak Integration Report (Sample 0309-3)

Signal: VWD1A, Wavelength=254 nm

Ret. Time [min]TypeWidth [min]AreaHeightArea %
8.676VB0.5532.752.980.33
9.482BV0.2851.925.690.52
9.661VV0.1729.283.850.29
9.856VB0.2441.115.720.41
10.465BV0.2990.3110.470.90
10.773VB0.3649.126.420.49
11.118BV0.2354.969.360.55
11.379VB0.2435.336.300.35
11.587BB0.1842.388.100.42
11.814BV0.26150.7915.721.50
12.014VV0.1799.1615.150.99
12.183VV0.19111.0713.311.10
12.449VB0.41193.9314.001.93
12.832BV0.38292.1925.822.90
13.130VB0.1821.703.090.22
13.666BB0.2323.423.150.23
15.835BM m2.082204.1661.9421.91
16.086MM m0.784406.36265.7043.80
17.272MM m2.072129.8158.4821.17
Totals10059.76100.00

HPLC Data – 0309-4 (PoresynSolutions’ 2000A Co-Polymer Coated CPG DMT-ON)

HPLC Injection Parameters – Sample 0309-4
HPLC Acquisition Parameters Sample ID: 0309-4
Data File0309-4 20260310 183403.dx
Sequence NameSingleSample
Project NameTest
Sample Name0309-4
Operatoradmin
InstrumentLC1260II
Injection Date2026-03-11 13:43:49+08:00
Injection Volume2.000 μL
LocationP2-D8
Acquisition MethodDMT-on.amx
TypeSample
Processing MethodDNA.pmx
Sample Amount0.00
Manual ModificationNone (无)

HPLC Peak Analysis Report – Sample 0309-4,Signal: VWD1A, Wavelength=254 nm

Ret. Time [min]TypeWidth [min]AreaHeightArea %
3.682BB1.121967.23252.586.94
6.108BB0.4523.802.180.08
6.722BV0.1314.713.310.05
6.813VB0.1521.924.250.08
7.095BV0.1322.765.760.08
7.235VB0.1973.4413.970.26
7.418BV0.0818.016.060.06
7.497VB0.1666.8012.240.24
7.674BV0.0923.157.500.08
7.737VB0.1246.4110.660.16
7.939BV0.1032.768.820.12
8.030VB0.1147.4410.970.17
8.192BV0.1573.8514.440.26
8.335VB0.1221.842.390.08
8.465BV0.17245.2943.370.87
8.959VV0.627403.59412.5626.13
9.282VB0.673764.71263.0813.29
10.028BV0.2545.845.040.16
10.221VB0.4678.953.270.28
12.343BV0.962465.34131.428.70
12.580VV1.1311853.501228.3541.84
13.543VB0.4019.672.380.07
Totals28331.03100.00